4 connecting to line power, 5 placement and rack mounting, Figure 1 pem (power entry module) – Fluke 1595A User Manual

Page 26

Advertising
background image

1594A/1595A Super-Thermometer

Connecting to Line Power

16

1. Disconnect line power.
2. Observe what line voltage setting is displayed in the PEM window. If it is correct, the fuse assembly

will be re-inserted in the same orientation it is removed. Otherwise, it will need to be rotated 180°

before re-insertion.

3. Open the fuse compartment by inserting a screwdriver blade in the slot located at the top of the fuse

compartment and open the PEM door.

4. Use the screwdriver blade to pry out the fuse block by inserting the blade in the slot located at the top of

the fuse block.

5. Remove the fuses from the assembly for replacement or verification. Be sure the correct fuses are

installed.

6. Reinstall the fuse assembly by pushing it back into the PEM while ensuring that the correct line voltage

label is shown in the PEM window. Close the PEM door so it locks in place.

Figure 1 PEM (Power Entry Module)

3.4 Connecting to Line Power

WARNING:

To avoid electrical shock, connect the factory supplied three-prong line power cord to a

properly grounded power outlet. Do not use a two-prong adapter or extension cord that will break the

ground connection.

After the line voltage setting and correct fuse are verified, connect the instrument to a properly grounded

three-prong outlet using the provided line voltage cord.

3.5 Placement and Rack Mounting

In general, place the Super-Thermometer in an area free of drafts and excessive electrical noise. Refer to the

specifications for environmental requirements.
The Super-Thermometer is designed to be used on a bench-top or installed in a standard width rack with the

optional rack mount kit. For bench-top use it is equipped with hinged, non-slip feet. For rack mount instruc-

tions, refer to the rack mount kit instruction manual.