1 scan settings (f1), 2 timing settings (f2), Figure 13 measurement process block diagram – Fluke 1595A User Manual

Page 49: Scan settings (f1), Timing settings (f2)

Advertising
background image

39

Menus and Screens

Main Menu

The following are complete descriptions of each function key.

5.7.3.1 SCAN SETTINGS (F1)

MAIN MENU | MEASURE MENU |

Scan Settings is used to set up channel scanning. The Super-Thermometer has several scan options that are

configured using the following fields.
SCAN ENABLE- turns scanning on or off by using the Left/Right Arrow Keys and pressing the ENTER key.

ON enables automatic sequencing through all enabled channels. OFF disables scanning so only one channel is

measured as selected by the user.
REPEAT COUNT- determines how many times a channel is measured before the Super-Thermometer contin-

ues to the next channel in the scan sequence. The setting has a range of 1 to 100.
ALTERNATE REFERENCE- automatically measures the reference channel after each individual channel is

measured in the scan sequence. This feature is useful in comparison type calibrations when the heat source is

not very stable.
REFERENCE CHANNEL- the setting that determines which channel is the reference channel for Alternate

Reference. The channel chosen for REFERENCE CHANNEL must be enabled (Section 5.7.1, CHANNEL

SETUP (F1), on page 29).

5.7.3.2 TIMING SETTINGS (F2)

MAIN MENU | MEASURE MENU |

The Timing Settings screen contains the settings and function keys needed for the configuration of measure-

ment timing. The timing settings control the measurement process and the length of time it takes to produce

and display a measurement.
Measurement Timing Description
The measurement timing settings allow the user to be in control of each phase of the measurement process.

These settings enable the user to optimize timing for different applications.
To help describe the measurement process, it is broken into the following five steps.

Figure 13

Step 1

Sampling

Defined by:

Sample Period

Step 2

Averaging

Enabled with:

Average Enable

Step 3

Filtering

Enabled with:

Digital Filter

Enable

Step 4

Conversion

Defined by:

Probe Conversion

Step 5

Display

Defined by:

Display Interval

Measurement Process Block Diagram

1. Sampling - the fundamental measurement of the Super-Thermometer is the resistance ratio. The

fundamental ratio measurement is sampled once per Sample Period.

2. Averaging - the fundamental ratio measurements can be averaged (see Average Enable below) or

passed directly to Filtering. The number averaged is the number of measurement samples taken during

one Display Interval (how often the measurement display is updated). For example, if Sample Period

is set to 2 seconds and Display Interval is set to 10 seconds, 5 ratio measurement samples are averaged

per Display Interval. The display is updated with the statistical mean of the 5 measurement samples.

If AVERAGE ENABLE is OFF, only the most recent ratio measurement sample will be passed to

Filtering.

3. Filtering - after Averaging, the measurement samples can also be processed with a digital filter

(configured by using the Digital Filter function key, Section 5.7.3.3, DIGITAL FILTER (F3), on page

Advertising
This manual is related to the following products: