5 temperature unit (f5), 5 system menu (f5), 1 time date (f1) – Fluke 1595A User Manual

Page 60: Figure 15 system menu screen, Temperature unit (f5), Time date (f1)

Advertising
background image

1594A/1595A Super-Thermometer

Main Menu

50

TRACE - controls whether a measurement point is plotted with a line that connects to the previous point or as

a separate dot. The options are LINE and DOT. The default value is DOT.

Important:

When Time Span is large, multiple points may be combined and drawn as a vertical line to

conserve memory.

5.7.4.5 TEMPERATURE UNIT (F5)

MAIN MENU | DISPLAY MENU |

The Temperature Unit screen allows the user to select the unit for displayed temperature measurements. The

options are °C, K, and °F. °F will not be available if SI Unit Lock is set to ON in the Calibration Parameters

screen of the Calibration menu (see Section 5.7.7.2, CALIBRATION (F5), on page 55). The default value is

°C.

5.7.5 SYSTEM MENU (F5)

MAIN MENU |

Figure 15 System Menu Screen

The System Menu contains general system settings such as time and date, remote interface, password, and

calibration.
The System Menu screen displays the following product information: Manufacturer, Model, Serial Number,

Firmware Version, and Calibration Date.
The following are the System Menu function keys:

5.7.5.1 TIME DATE (F1)

MAIN MENU | SYSTEM MENU |

The Time and Date function key allows the user to enter the system Time and Date using the following fields:
TIME - time of the system clock. It is always in 24-hour format.
DATE - date of the system clock. It appears in the format determined by the Date Format setting (see Section

5.7.4.1, USER SETTINGS (F1), on page 47).