4 *idn, 5 *rst, 6 *sre <value – Fluke 1595A User Manual

Page 85: 4 *idn? 8.7.3.5 *rst 8.7.3.6, Sre <value, Remote operation, Any self-test in progress is canceled, Measurement enable is set to off, Scan enable is set to on, Repeat count is set to 1

Advertising
background image

75

Remote Operation

Interface Commands

Bit

Event

0

OPC, Operation Complete

1

Request control (not used)

2

Query error

3

Device dependent error

4

Execution error

5

Command error

6

User request (not used)

7

Power on

8.7.3.4 *IDN?

Example response: FLUKE,1594,A9C111,1.10/1.00
Reads the instrument information.
The response is given in the following format:
<manufacturer>,<model number>,<serial number>,<version numbers>

8.7.3.5 *RST

Sets operating settings to default states.
The following settings are affected:

Any self-test in progress is canceled

Measurement enable is set to OFF

Channel 2 enable is set to ON, all other channels OFF

Scan Enable is set to ON

Repeat Count is set to 1

Alternate Reference is set to OFF

Reference Channel is set to channel 2

Timing settings are preset to NORMAL presets

Stop Timer Enable is set to OFF

Timer Duration is set to 3600 s

Timer Beep is set to ON

Filter Enable is set to ON

Time Window is set to 60

Statistics Mode is set to UNLIMITED

Sample Count is set to 100

Delta Mode is set to OFFSET

Delta Offset is set to 0.

Delta Channel is set to channel 2

Unit is set to °C

The key lock setting is set to off

Statistics are reset

Not affected are channel settings other than enable, probe definitions, resistor definitions, recording settings,

display settings other than Unit, remote interface settings, time and date, password settings, and calibration

settings.

8.7.3.6 *SRE <value>

Example command: *SRE 16