8.7.3.29 DATA:REC:FILE <name>

Query form: DATA:REC:FILE?
Sets the data recording USB File name (see Section 5.7.3.4.1, RECORDING SETTINGS (F1), on page 42).

8.7.3.30 DATA:REC:MEM INT|USB|BOTH

Query form: DATA:REC:MEM?
Sets the data recording Memory Device (see Section 5.7.3.4.1, RECORDING SETTINGS (F1), on page

42).

8.7.3.31 DATA:REC[:STAT] <boolean>

Query form: DATA:REC?
Starts or stops recording. This is equivalent to pressing the START RECORDING or STOP RECORDING

function key in the Main Screen. A <boolean> parameter value 1 enables recording; 0 disables recording.
The *RST command has no effect on this setting.

8.7.3.32 DISP:ALER:CLEA

Clears any alert message that is active.

8.7.3.33 DISP:ALER[:READ]?

Example response: 1,“Maximum temperature exceeded”
Reads the active alert message. The response given is the alert message number and description, comma-sepa-

rated and with the description enclosed in quotation marks. If no alert message is active, the response is 0,”No

alert”. The following table lists the possible responses:

Alert

Cause

0,“No alert”

No alert message is active.

1,“Maximum temperature exceeded”

A measured temperature exceeds the limit defined for a probe.

2,“ITS-90 subrange exceeded”

A measured temperature is outside the ITS-90 subranges defined for
a probe.

3,“Resistor due for calibration”

A defined resistor in the Resistor Library is past due for calibration.

The active alert message may be cleared using the DISP:ALER:CLEA command.

8.7.3.34 DISP:DATE:FORM YMD|DMY|MDY

Query form: DISP:DATE:FORM?
Alternate parameter: DEF
Sets the display Date Format setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default

setting is YMD.

8.7.3.35 DISP:DEC:FORM PER|COMM

Query form: DISP:DATE:FORM?
Alternate parameter: DEF
Sets the display Decimal setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default set-

ting is PER.

8.7.3.36 DISP:FIEL:FEED<n>?

Example command: DISP:FIEL:FEED4?
Example response: DEF