59 inp:prob:par <id>,<name>,<value, Inp:prob:par <id>,<name>,<value – Fluke 1595A User Manual
Page 96
1594A/1595A Super-Thermometer
Interface Commands
86
Reads the ID of the next probe in the Probe Library. It also advances the library index used for the next re-
sponse to the INP:PROB:NEXT? command.
If there are no more probes in the Probe Library, the response is an empty string, “”.
8.7.3.59 INP:PROB:PAR <id>,<name>,<value>
Example command: INP:PROB:PAR “PRT_A46002”,TYPE,PRT
Query form: INP:PROB:PAR? <id>,<name>
Example query command: INP:PROB:PAR? “PRT_A46002”,TYPE
Example query response: PRT
Password protection: Probe Protection
Sets the value of a setting in a probe definition.
Parameter <id> is the ID of a probe definition in the Probe Library. The ID should be enclosed in double
quotation marks. If <id> does not match the ID of a probe definition in the Probe Library, error -224,“Illegal
parameter value” is generated in the error queue.
Parameter <name> specifies the setting to set. Settings and their associated names are listed in the table below
(also see Section 5.7.2.3, EDIT PROBE (F3), on page 36).
Setting
<name>
<value> type or options
Type
TYPE
PRT|THER|RES
Calibration Date
CAL_DATE
string: “<yyyy>-<mm>-<dd>”
Nominal Current
NOM_CURR
decimal number
Maximum Current
MAX_CURR
decimal number
PRT type settings
Maximum Temperature
MAX_TEMP
decimal number, dependent on Unit setting
Conversion
CONV
ITS_90|PT_100|CVD_ABC|CVD_ALPH|POLY|NONE
ITS-90 conversion settings
RTPW
RTPW
decimal number
High Subrange
HIGH_SUB
NONE|5|6|7|8|9|10|11
Low Subrange
LOW_SUB
NONE|1|2|3|4|5
ITS-90 subrange settings
A1 – A11
A1 – A11
exponential number
B1 – B9
B1 – B9
exponential number
C1 – C5, C6, C7
C1 – C5, C6, C7
exponential number
D
D
exponential number
PT-100 conversion settings
R0
R0
decimal number
CVD-ABC conversion settings
R0
R0
decimal number
A
A
exponential number
B
B
exponential number
C
C
exponential number
CVD-ALPHA conversion settings
R0
R0
decimal number
ALPHA
ALPH
exponential number
DELTA
DELT
exponential number
BETA
BETA
exponential number
Polynomial conversion settings