Field of view, Emissivity, Hold – Fluke 566 User Manual

Page 22

Advertising
background image

566/568

Users Manual

14

Field of View

For accurate measurements, make sure that the target is larger than the unit’s spot size.

The smaller the target, the closer you should be to it (see Figure 6).

Yes

N

o

eyl05.eps

Figure 6. Field of View

Emissivity

Emissivity describes the energy-emitting characteristics of materials. Most organic

materials and painted or oxidized surfaces have an emissivity of approximately 0.95, the

default setting for the thermometer.
To compensate for inaccurate readings that may result from measuring shiny metal

surfaces, you can cover the surface to be measured with electrical tape or flat black paint

(<148

°

C/300

°

F) with emissivity set to 0.95. Allow time for the tape or paint to reach the

same temperature as the surface beneath it. Measure the temperature of the tape or

painted surface.
If you cannot paint or use tape, then you can improve the accuracy of your measurements

by either numerically adjusting emissivity or by using the Emissivity Menu to access a

table of some common materials. The thermometers have numerically adjustable

emissivity from 0.10 to 1.00 that can be used in conjunction with the emissivities in

Table 3. Also, the thermometers have a built-in table of the most common materials that

can be set from the Emissivity Menu. The built-in values are indicated in Table 3.

HOLD

The display retains its last infrared measurement for twenty seconds when the trigger is

released, with HOLD appearing on the display. With the probe inserted, the contact

thermometer remains on. To freeze the infrared temperature when a probe is not inserted,

release the trigger until HOLD appears on the display.

Advertising