Example 1:

A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients Rtpw, a4, b4, a8, and b8. Set the 1502A parameters with values
from the certificate as follows.

Example 2:

A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients Rtpw, a5 and b5. Set the 1502A parameters with values from the
certificate as follows:

19

6 General Operation

Selecting the Probe Characterization

1502A Coefficient

Certificate Value

R0.01

Rtpw

A

a8

b

b8

C

0

d

0

A4

a4

b4

b4

Table 3 Setting Coefficients Rtpw, a8, b8, a4, and b4

1502A Coefficient

Certificate Value

R0.01

Rtpw

A

a5

b

b5

C

0

d

0

A4

a5

b4

b5

Table 4 Setting Coefficients Rtpw, a5, and b5