Table 6 command list 27, Table 6, Command list – Fluke HART SCIENTIFIC 1502A User Manual
Page 31
27
7 Digital Communications Interface
Remote Commands
Command
Description
Measurement Commands
T
read measurement (includes label, unit, and time)
F[ETCH?]
read measurement value (SCPI compatible)
SA[=[[[hh:]mm:]ss]
read [or set] serial sample period
U=C|F|K|O
select units
ST[=ON/OF]
read [or set] the time stamp
CL[=hh:mm:ss]
read [or set] the system clock
Probe Characterization Commands
PR[=90|68|R|S]
read [or select] the characterization type
R0[=<value>]
read [or set] R0 or R(0.01°C) depending on the selected characterization
AL[=<value>]
read [or set]
α for the Callendar-Van Dusen or IPTS-68 characterization
DE[=<value>]
read [or set]
δ for the Callendar-Van Dusen or IPTS-68 characterization
BE[=<value>]
red [or set]
β for the Callendar-Van Dusen characterization
A4[=<value>]
read [or set]
a
4
for the ITS-90 or IPTS-68 characterization
B4[=<value>]
read [or set]
b
4
for the ITS-90 characterization
C4[=<value>]
read [or set]
c
4
for the IPTS-68 characterization
A
n[=<value>]
read [or set]
a
6
, a
7
, a
8
, a
9
, a
10
, or
a
11
for the ITS-90 characterization
B
n[=<value>]
read [or set]
b
6
, b
7
, b
8
, or b
9
for the ITS-90 characterization
C
n[=<value>]
read [or set]
c
6
or
c
7
for the ITS-90 characterization
D6[=<value>]
read [or set]
d
6
for the ITS-90 characterization
SC[=68/90]
read [or set] IPTS-68 scale conversion
CO=<value>
test resistance to temperature conversion
Sample Parameter Commands
FI[=<value>]
read [or set] filter time constant
CU[=<value>]
read [or set] probe current
PS[=<value.]
read [or set] the power saver period
Communication parameter commands
DU[=F/H]
read [or set] serial sample duplex mode
LF[=ON/OF]
read [or set] serial linefeed
Table 6 Command List