2 specifications – Fluke VICTOREEN 05-437 User Manual

Page 6

Advertising
background image

Victoreen 05-437
Operators Manual

1-2

1.2 Specifications

Detector Energy-compensated

G-M

tube


Energy Dependence

10% to 40% from 50 KeV to 2 MeV, as referenced to

137

Cs


Overload Protection

Alarms in fields higher than 100 R/h when detector is completely blocked for
pulse operation.

Operation Indicator

Provide continuous visual indication of background radiation.


Range/lndicators 1

mR/h:

Green

2

mR/h:

Green

4

mR/h:

Yellow

8

mR/h:

Yellow

16

mR/h:

Red

32 mR/h: Red

Response Time Level Response

Time

1 mR/h

3 seconds

2 mR/h

1.5 seconds

4 mR/h

800 milliseconds

8 mR/h

400 milliseconds

16 mR/h 200 milliseconds

32 mR/h 100 milliseconds


Alarms

Alarm may be preset by user at any indicated discrete level by means of the
front panel selector switch; front panel switch selects all alarms or visual
only.

Visual Alarm: bright incandescent red lamps flashing on/off approximately
1/sec.

Aural Alarm: 400 Hz tone, alternating 1/sec; sound level = 76 db at one
foot.


Operating Conditions

10° C to 40° C (50° F to 104° F)

Maximum of 90% relative humidity (non-condensing)


External Connectors

Jack located on case bottom allows attachment

of external aural/visual

alarm (Primalarm) at remote location.


Power Supply

Part 14-314 (United States, Canada)

Input 117 VAC, 60 Hz, 12 W

Output 12 VDC, 500 mA


Part 14-400 (Europe)

Input 230 VAC, 50 Hz,

Output 12 VDC, 500 mA


Part 14-417 (United Kingdom)