Initial Start-Up

3.1

Connection to the Mains (230 V, 50 Hz)

➭ Connect the mains plug at the test instrument to the mains power

outlet. The function selector switch can be set to any position.
If a mains outlet (earthing contact outlet) is not available, or if only a
3-phase outlet is available, the adapter socket can be used to connect
the phase conductor, the neutral conductor and the protective
conductor. The adapter socket has three permanently attached cables
and is included with the KS13 cable set

Attention!

!

If connection is not possible via an earthing contact outlet: Shut
down mains power first.
Then connect the cables from the coupling socket to the mains
using pick-off clips in accordance with the diagram.
Disconnection from mains power is only possible with the mains
plug.

L1

N

green-yellow

green-yellow

PE

L1

L2

L3

N

PE

L1

L2

L3

N

green-yellow

U

L – N

= 230 V

Mains Connection