GE F-801 User Manual

Page 50

Advertising
background image

F-801 Internal RO System by GE Osmonics

12552e – 27 Feb 04

4- 10

Routine Maintenance

4.

Carefully remove the housing (bottom part) from its head. Empty the water and
used pre-filter cartridge.

5.

Inspect the central core of the used pre-filter for dirt or rust. If discoloration is
present in the cartridge interior, the filter has been in service too long and
should be replaced sooner, or at a delta pressure lower than 8 PSI.

6.

Check the O-ring in the groove of the filter housing. Replace the O-ring (GE
Osmonics p/n 40694) if it appears worn or damaged.

7.

Clean and rinse the pre-filter housing.
a.

Optional; a 1% bleach solution may be used to sanitize the housing, rinse
thoroughly.

8.

Insert the new pre-filter cartridge into the housing. Ensure that the pre-filter is
firmly in place.

9.

Re-install the housing, tighten clockwise.

10.

Turn ON the feed water and check for leaks. This will also purge the air from
the pre-filter housing.

11.

If there are no leaks, the RO system is ready for use.

Advertising