Tubing Care:

Inspect tubing after each pumping session for

condensation and or milk.

If condensation appears in tubing:

To eliminate condensation after you have completed

pumping, continue running the pump with the tubing(s)
attached for another 1-2 minutes or until dry, while you
attend to storing your pumped breastmilk.

If milk appears in tubing:

1. Turn off the pump and unplug from power source.

2. Remove and disassemble the tubing, protective mem-

brane and membrane cap. Wash in soapy water then
rinse in cold clear water.

3. Shake water droplets out and hang to air-dry. For

faster drying, attach tubings to pump and run pump 1-2
minutes or until dry, or pour a small amount of iso-
propyl alcohol through the tubings to dry.

4. Make sure that valve membranes are undamaged and

clean. Replace if necessary.

5. Use a damp cloth to wipe the pump and the

diaphragm. Do not immerse the pump in water.

6. When components are completely dry, reassemble

breastpump and kit.

Important Battery Care

Charge Symphony

®

pump batteries for 12 hours upon receipt.

Pump may be used while charging.

Optimal battery life is achieved when:

Batteries are not fully drained on a regular basis

Pump is allowed to charge when not in use

If unable to charge regularly, charge for 12 hours
every 2 months at minimum

Pumping time with a completely charged battery is
approximately 60 minutes.
4

1.

PersonalFit

Breastshield, Standard (24mm)....................87073

2.

PersonalFit Breastshield Connector ................................87071

3.

Valve & Membrane ........................................................8000001

4.

Collection Connector ....................................................8117009

5.

Solid Container Lid ......................................................8107057

6.

Tubing............................................................................8107168

7.

Membrane Cap ............................................................8100769

8.

Protective Membrane....................................................8200094

9.

Breastfeeding Information Guide ..................................1907200

10. Symphony Instruction Manual ......................................1907724