MaxTech MAXO2 OM-25ME User Manual

Page 16

Advertising
background image

6) Use the

↑ and ↓ arrow keys to adjust the displayed oxygen

concentration to the level of the known concentration. Pressing the
arrow keys changes the value in .1% increments. If the keys are held
down for more than 1 second the display will scroll at a rate of .4%
per second.

Note: If 10 seconds elapse between key actuations, the system will
store the latest calibration value and will revert to normal operation.
If this occurs inadvertently, simply repeat the calibration procedure.

7) When the calibration value is set, press the CALIBRATE or

LOCK\UNLOCK key again to accept the calibration setting and
return to normal operation.

Note: If the message "CAL," followed by the message "Er" flashes
on the display after entering the desired calibration value, the
system has determined that the entered value will not allow
operation within the specified output range of the sensor. This
situation may occur if:

a) the operator has inadvertently entered the wrong concentration

for the calibration gas.

b) the concentration of the calibration gas is not correct.

c) the sensor is in need of replacement.

d) the operator attempted to adjust the monitor before allowing

sufficient time for the calibration gas to purge out the previous
sample.

e) the flow and pressure of the calibration gas was not properly

regulated.

Check these items and repeat calibration. If calibration error
continues to occur, contact the service department of the distributor
from which the unit was purchased, or you may call Maxtec's
Customer Service Department directly.

2.2.3 Automatic Calibration to Room Air

The MAXO

2

Monitor can quickly be calibrated to room air (20.9%) using

a quick-key shortcut command. This function saves time by setting the
calibration value to 20.9% without scrolling the display.

9

R213M05H.qxd 6/26/2006 9:04 AM Page 8

Advertising