MaxTech MAXO2 OM-25ME User Manual

Page 23

Advertising
background image

16

7. SPECIFICATIONS

7.1 Base Unit Specifications

Measurement Range:

0.0-100%

Resolution:

0.1%

Accuracy and Linearity: ±1% of full scale at constant temperature, R.H.

and pressure when calibrated at full scale.

Total Accuracy:

±3% Actual Oxygen Level over full operating
temperature range.

Response Time:

90% of final value in approximately 15 seconds
at 23 C

Warm-up Time:

none required

Operating Temperature: 15 C - 40 C (59 F - 104 F)
Storage Temperature:

-15 C - 50 C (5 F - 122 F)

Humidity:

0-95% (non-condensing)

Power Requirements:

2, AA Alkaline batteries (2 X 1.5 Volts)

Battery Life:

approximately 3000 hours in typical use.

Low Battery Indication: "LOW BAT" icon displayed on LCD
Sensor Type:

Maxtec MAX-250E galvanic fuel cell

Expected Sensor Life:

>900,000% O

2

Hours

over 2 years in typical applications

Alarm System:

high/low alarms, flashing red LEDs, nominal
2300Hz audible alarm beeper

Low Alarm Range:

18%-99% (>1% lower than high alarm)
15%-99% (>1% lower than high alarm)OM-25ME-L

High Alarm Range:

19%-99% (>1% higher than low alarm)
16%-99% (>1% higher than low alarm)OM-25ME-L

Alarm Accuracy:

exact to displayed alarm value

Dimensions:

3.5"(W) x 5.5"(H) x 1.5"(D) [89mm x 140mm x 38mm]

Weight:

approximately .92 lbs. (417g)

Cable Length:

10 ft. (3m) fully extended

Diverter Fitting:

fits industry standard, 15 mm "T" adapter

7.2 Sensor Specifications

Type:

galvanic fuel sensor (0-100%)

Life:

2 years in typical applications

Interferent

Volume % Dry

Interference in O

2

%

Nitrous Oxide

75%

<2%

Carbon Dioxide

10%

<2%

Halothane

5%

<2%

Enflurane

5%

<2%

Isoflurane

5%

<2%

Helium

70%

<2%

Sevoflurane

6%

<2%

Desflurane

15%

<2%