Soak in hot water/vinegar solution for 30 minutes, Changing the nebulizer, The nebulizer is designed for single patient use – Invacare IRC 1720 User Manual

Page 30: Cambio de nebulizador, Changement de nébuliseur

Advertising
background image

SECTION 4—CLEANING AND DISINFECTING

Stratos™Portable

30

Part No 1130202

4. Thoroughly clean the nebulizer and mouthpiece in hot

water/detergent solution.

5. Remove from the water/detergent solution and rinse with

clear hot tap water.

6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.
NOTE: If using medical disinfectant cleaners, follow

manufacturer’s instructions carefully.
8. Keep the outer surface of the tubing dust-free by wiping

regularly.

NOTE: The nebulizer air tubing does not have to be washed

internally because only filtered air passes through the tubing

Changing the Nebulizer

The nebulizer is designed for single patient use.
The nebulizer must be replaced after a long period of

inactivity, or when it is obstructed by dry medication, dust,

etc. The reusable nebulizer should last from six months to one

year with general use. The disposable nebulizer should last 30

days.