SECTION 6—TROUBLESHOOTING

Stratos™Portable

36

Part No 1130202

Nebulizer does not work,

and the light indicator is red.

The thermal protector has been activated due

to:

a. The device operating outside the conditions

outlined in this manual.

b. The device has been operating near heat

sources or in hot environments.

Allow the nebulizer to cool down for a mini-

mum of 40 minutes. Wait until the RED light

turns off before turning the nebulizer on again.

Optional battery is not

working optimally.

Battery was not properly conditioned (charged

for at least 16 hours before first use).

If it has been in use for approximately four

months, it may be time to recondition the bat-

tery.

Optional battery is not

charging.

Battery will not charge if connected to the nebu-

lizer.

Check all power cord connections and cord

conditions.

Check power outlet.

PROBLEM

SOLUTION