Operating the scale, Weighing the patient, Keypad functions – Invacare Jasmine JLS5 User Manual

Page 4: Ƽ warning

Advertising
background image

Jasmine™ Scale JLS5

4

Part No 1152408

Operating the Scale

Keypad Functions

FIGURE 3 Operating the Scale - Keypad Functions

Weighing the Patient

ƽ WARNING

The weight capacity is limited to the lowest rated capacity of any one of the components in use

(e.g. Patient Lift, Sling or Scale). The patient's weight MUST NOT exceed the lowest rated

capacity of any component.

1. Attach sling straps to the swivel bar. For proper attachment instructions refer to Attaching a Sling 

to the Swivel Bar in the Jasmine Patient Lift Owner’s Manual, Invacare P/N 1150704.

NOTE: For improved accuracy on the patient lift, attach all sling straps to the center hooks on both sides of the 
swivel bar. However, patient comfort may require a different configuration of sling straps.

2.  Press the ON/OFF key.

KEY

INDICATOR

DISPLAYED

INDICATOR

LOCATION

DEFINITION

ON

OFF

OFF

Center of Display Window

Pressing this key will apply power to the
scale and turn the unit on. When the scale
is already on, pressing the button will turn
the unit off.

ZERO

ZERO

Lower Left Corner of the
Display Window

Pressing this key when the scale is on will
reset the weight shown in the display
window to zero.

UNITS

lb or kg

Upper Right Corner of the
Display Window

The function of this key is to change the
unit of measurement from pounds (lb) to
kilograms (kg).

LOCK

UNLOCK

LOCK

Lower Right Corner of the Display
Window

This key is used to LOCK or UNLOCK a
weight value in the display window.
NOTE: Weight displayed is a STORED
weight and not the current weight when
the key is used in the lock position.

N/A

LO BAT

Center of Display Window

Indicator is shown in the display window
to notify when battery is low.

Jasmine Scale

JLS5

CAPACITY 500 LB, 227 KG

Operation

Keys

Display

Window

Advertising