Replacing the battery, Calibrate the jasmine scale – Invacare Jasmine JLS5 User Manual

Page 6

Advertising
background image

Jasmine™ Scale JLS5

6

Part No 1152408

Replacing the Battery

NOTE: For this procedure, refer to FIGURE 4.

NOTE: The scale is powered by a 9‐volt alkaline battery that should provide approximately 1500 readings before 
needing replacement.

When battery replacement is needed, LO BAT will appear on the display. Perform the following:

1. Slide the battery door open in the direction of the arrow.

2. Remove existing battery.

3. Install the NEW battery.

4. Reinstall the battery door.

FIGURE 4 Replacing the Battery

Calibrate the Jasmine Scale

NOTE: For this procedure, refer to FIGURE 5 on page 7.

NOTE: The Jasmine Scale will be pre‐calibrated at the factory with the load cell. Should it be necessary to re‐calibrate 
the scale, follow the instructions outlined below.

1. The patient and the sling MUST be removed from the scale to properly calibrate the Jasmine Scale. Refer 

to Operation in the Jasmine Patient Lift Owner’s Manual, Invacare P/N 1150704, for further instructions.

2. For sling detachment instructions, refer to Lifting the Patient in the Jasmine Patient Lift Owner’s Manual, 

Invacare P/N 1150704.

3. Turn the scale display power on by pressing the ON/OFF key.

4. Rotate the Swivel Bar until the back of the scale display is visible.

5. Insert the tip of a paper clip or similar tool into the small hole on the back of the scale display and press 

the CAL button.

NOTE: Once the CAL button is pressed, “CAL1” will be seen in the display window.

6. Press the UNITS key. The display reads CAL2.

7. Press the LOCK/UNLOCK key. The display reads UNLD.

8. Press the ZERO key. The display reads LOAD.

9. Hang 200 pounds of calibrated weight from the scale and press the ZERO key.

10. Press the ZERO key. The display reads DONE.

Battery

Door

9V Battery

Assembled

Scale

Arrow

Advertising