Oxygen purity, Senso, Sensor feature – Invacare IRC5LXO2 User Manual

Page 16

Advertising
background image

16

INITIAL STARTUP OF THE CONCENTRATOR

NOTE: Concentrator may be used during the initial start warm-up time

(approximately 30 min.) while waiting for the GREEN O

2

purity light to illuminate.

When the unit is initially turned on, the GREEN power LED (I/O) will

illuminate. Within 5 to 30 minutes after initial startup, a GREEN light will

illuminate in the Oxygen Purity section of the Control Panel (FIGURE

7) if concentration is within specification. Any other SensO

2

® indicator lights

illuminating will signal a condition that requires a call to the supplier.

OPERATING INSTRUCTIONS

EXPLANATION OF OXYGEN PURITY INDICATOR

LIGHTS (FIGURE 7):

GREEN LIGHT ( ) - Normal Operation
YELLOW LIGHT ( ) - Call Supplier IMMEDIATELY. You may

continue to use the concentrator unless instructed otherwise by your

supplier. Be certain that BACKUP OXYGEN is nearby.
RED LIGHT ( )- Total unit shutdown. Switch IMMEDIATELY

to BACKUP OXYGEN supply. Call Supplier IMMEDIATELY.
GREEN LIGHT - with YELLOW LIGHT FLASHING - Call

Supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may

continue to use the concentrator.

O

2

FIGURE 7 - O

2

DISPLAY CONTROL PANEL

Call Supplier

Normal

Below Normal

Oxygen Purity

RED

YELLOW

GREEN

O

2

Auto Shut - Down

Use Backup

SENSO

2

® OXYGEN PURITY INDICATOR-

IF YOUR UNIT FEATURES THE O

2

SENSOR

FEATURE

This feature monitors the purity of oxygen generated by the oxygen

concentrator. If purity falls below factory preset standards, indicator

lights on the control panel will illuminate.