Specifications – Invacare IRC5LXO2 User Manual

Page 8

Advertising
background image

8

SPECIFICATIONS

Electrical Requirements:

Rated Current Input:

Sound Level:

Altitude:

*Oxygen Output

Concentration Levels:

Maximum Outlet Pressure:

120 VAC +10% (132 VAC/108 VAC),

60 Hz

4.0 A

50 dB Average

Up to 6,000 ft. (1828 meters) above

sea level without degradation of

concentration levels. From 6,000 ft.

(1828 meters) to 13,129 ft. (4000

meters) below 90% efficiency.

ALL 5LXO2/5LX MODELS

95.6% to 87% at 1/2 to 5 L/min.

* Stated concentration levels

achieved after initial warm-up

period (approximately 30 minutes).

5 + 0.5 psi.

(34.5 + 3.45 kPa.)

SPECIFICATIONS

SensO

2

® Alarm Thresholds:

O2 Purity

Below 73%

73% to 85%

Over 85%

Indicators

RED Indicator Light -

Continuous Audible Alarm

Sieve-GARD™Compressor

Shutdown

YELLOW Indicator Light

GREEN Indicator Light

O

2

Label

Symbol

Advertising