SECTION 6—MAINTENANCE

Perfecto

2

Series

26

Part No 1143482

FIGURE 6.1 Cleaning the Cabinet Filter

Cabinet

ƽ WARNING

UNPLUG the concentrator when cleaning. To avoid

electrical shock, DO NOT remove cabinet.

1. Clean the cabinet with a mild household cleaner and

non-abrasive cloth or sponge.

Humidifier

NOTE: To clean the oxygen humidifier, follow the instructions
provided by the manufacturer. If none are provided, follow these
STEPS:

1. Clean the humidifier every day.

2. Wash it in soapy water and rinse it with a solution of ten

parts water and one part vinegar.

3. Rinse thoroughly with hot water and refill with distilled

water to the level shown on the humidifier.

ƽ WARNING

DO NOT overfill humidifier bottle.