SECTION 4—TYPICAL PRODUCT PARAMETERS

Part No 1143482

13

Perfecto

2

Series

SECTION 4—TYPICAL PRODUCT
PARAMETERS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60
Hz

Rated Current Input:

Perfecto

2

5: 3 A

Sound Level:

Perfecto

2

5: 44 ± 2 dBA Average

Altitude:

Perfecto

2

5 - Up to 6,000 ft. (1828 meters)

above sea level without degradation of con-
centration levels. From 6,000 ft. (1828
meters) to 13,129 ft. (4000 meters) below
90% efficiency.

*Oxygen Output
Concentration Levels:

All IRC5PO2/IRC5P models 95.6% to 87%
at 0.5 to 5 L/min

Maximum Outlet Pressure:

Perfecto

2

5: 5 psi ± 0.5 psi. (34.5 kPa ± 3.45

kPa.)

Flow Range:

Perfecto

2

5 - 0.5 to 5 L/min

For flowrates less than 1 L/min, we recom-
mend the use of the Invacare Pediatric Flow-
meter IRCPF16 Accessory.

Low Flow Alarm:

Perfecto

2

5 - 0 L/min to 0.5 L/min

Rapid Audible Alarm Beeping (No
Accessories Connected).

Average Power
Consumption:

Perfecto

2

5: 280 W (Operating @ 3 L/min)

Pressure Relief Mechanism
Operational at:

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)

Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:

0.7 L/min

Filters:

Cabinet, Outlet HEPA and Compressor
Inlet