Supplemental information, Compressor, Low voltage can also be suspected when – Invacare IRC 1705 User Manual

Page 14: The motor does not get up to full power or speed, Extension cords

Advertising
background image

SUPPLEMENTAL INFORMATION

Invacare

®

Select

14

Part No 1148073

SUPPLEMENTAL

INFORMATION

Compressor

If the unit shuts down frequently, you may have a low voltage 
situation. 

Low voltage can also be suspected when:
• The motor does not get up to full power or speed.
• Home fuses or circuit breakers activate when starting 

compressor.

• Lights dim or remain dim when compressor is started.
• Other motor operated appliances fail to operate properly 

or too many motor operated appliances are on same 
circuit.

Extension Cords

If you must use an extension cord, use only a cord having the 
same or higher electrical rating as the device being connected. 
Make certain the extension cord is in good condition without 
any exposed wires or hazardous contacts, and the extension 
cord wire is NOT LESS than 18 gauge for 25 feet.