SECTION 4—MAINTENANCE

Part No 1148073

25

Invacare

®

Select

Nebulizer Change

The nebulizer is designed for single patient use. The nebulizer 
MUST be replaced after a long period of inactivity, or when it 
is obstructed by dry medication, dust, etc. Reusable 
nebulizers should last from six months to one year with 
general use. Refer to the nebulizers owner’s manual. 
Disposable nebulizers should last 30 days. For disposable 
nebulizers, it is recommended that an extra nebulizer be kept 
at all times. Refer to the disposable nebulizer’s owner’s 
manual.

Filter Change

1. The filter MUST be replaced after approximately 30 hours 

of use or when it turns grey.

2. Open the filter holder to replace the filter with a new one.

CAUTION

DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate unit without an air filter.