Troubleshooting

Image Problems

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123

Probe Problems

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124

Power Problems

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124

Lamp Problems

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124

Software Problems

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125

For measurement problems, see

“Troubleshooting Inaccurate Measurements”

on

page 109.

Image Problems

Condition

Causes

Actions

Images are distorted or
blurred within normal
range.

• Optical tip is not

threaded onto the
camera head
securely.

• The optical surfaces

are dirty.

• Reinstall the optical tip. See

“To Attach an Optical Tip”

on page 31.

• Clean the optical tip and the camera head. See

“To Inspect and Clean an Optical Tip”

on

page 111 and

“To Inspect and Clean a Probe”

on page 112.

Image quality is
otherwise poor.

Various

• Verify that the QuickChange probe is properly seated and latched. See

“To Install a

Probe”

on page 30.

• Clean the optical tip and camera head. See

“To Inspect and Clean an Optical Tip”

on

page 111 and

“To Inspect and Clean a Probe”

on page 112.

• If the problem occurs in DC operation, check battery’s charge and its compliance with

listed specifications. For an illustration of the battery charge indicators, see

“Maintaining

the Battery”

on page 118.

• Verify that the lamp has been installed properly. See

“To Install a Lamp”

on page 115.

No image appears in
the display. (Green
LED on the power
switch is lit.)

Various

• Verify that the probe is properly seated and latched. See

“To Install a Probe”

on

page 30.

• Point the probe tip toward a bright light and see if image appears. If image appears, the

problem is associated with lamp operation. If still no image, contact GE Inspection
Technologies.

• If an external monitor is available, determine whether an image appears on it. Connect

the monitor’s VIDEO IN (or S-VIDEO IN) to the base unit’s VIDEO OUT (or S-VIDEO
OUT) connector. Verify the image.

Frozen image “jitters.”

Probe tip was moving
when image was
captured.

Unfreeze then refreeze the image, keeping probe tip steady.