Important safeguards/notices for use in the
medical environments

1. All the equipment connected to this unit shall be

certified according to Standard IEC60601-1,
IEC60950-1, IEC60065 or other IEC/ISO Standards
applicable to the equipment.

2. When this unit is used together with other equipment

in the patient area*, the equipment shall be either
powered by an isolation transformer or connected via
an additional protective earth terminal to system
ground unless it is certified according to Standard
IEC60601-1 and IEC60601-1-1.

3. The leakage current could increase when connected

to other equipment.

4. The operator should take precautions to avoid

touching the rear panel input and output circuitry and
the patient at the same time.

5. Model LMD-2140MD is a monitor intended for use

in a medical environment to display pictures from
cameras or other systems. These models are
nonpatient care equipment with respect to the
requirement of leakage current.

Caution

When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.

Symbols on the unit

R1.5m

* Patient Area

Symbol

Location

This symbol indicates

Bottom

Main power switch.
Press to turn the monitor
on.

Bottom

Main power switch.
Press to turn the monitor
off.

Bottom

The equipotential
terminal which brings the
various parts of a system
to the same potential.

Bottom

Functional earth terminal

8

Front

Brightness control

6

Front

Contrast control

-

Front

Key inhibit
The setting are locked so
that they cannot be
changed.

Warning on power connection
Use a proper power cord for your local power supply.

The United States

Canada

Continental Europe

Plug type

HOSPITAL GRADE

E41395

E41395-A

HOSPITAL GRADE

LL33182

LL76662

LP-34A

Female end

LS-60

Cord type

H05VV-F

Rated Voltage & Current

10A/125V

10A/125V

10A/250V

Safety approval

UL

CSA

VDE