Sunbeam 7659 User Manual

Page 6

Advertising
background image

Display mark

Condition/Cause

Corrective action

Mark

in the

Measurement in progress

surement condition and

remain quiet.

Measurement in

flashes when pulse is

progress

detected.

Mark flashes when power is

Automatic exhaust is

applied and there is air

formed.

remaining in the cuff.

Exhaust

-

-

A

Measurement has begun

Raise

but the pressure was

pressure 30-40

Insufficient

insufficient.

more and measure once

pressure

again.

Replace

batteries

Appears when the battery

Replace all four batteries

voltage is excessively tow.

with new ones.

error

Pulse

error

Appears when the blood
pressure could not be
obtained accurately.

Appears when the pulse
rate could not be obtained
accurately.

Exhaust the air from the
cuff, and re-apply the cuff
properly and re-measure.

(Wait

minutes before

re-measure)