Troubleshooting, Table 1 – Shoprider PHFW-1018 User Manual
Page 17
7/18/2011
17
Troubleshooting
The Self-Help Guide (Table 1) is intended to assist in the location of a
fault that may occur in a certain part of the power chair.
If after checking out the fault from the table below and the fault is still
showing, do not use the power chair. Turn off the power and consult
your provider immediately.
Table 1
BATTERY GAUGE
POSSIBLE FAULT
10 Bars flash
An excessive voltage has been applied to the control
system. This is usually caused by a poor battery
connection.
9 Bars flash
The parking brakes have a bad connection. Check the
parking brake and motor connections.
8 Bars flash
A control system fault is indicated. Make sure that all
connections are secure.
7 Bars flash
A joystick fault is indicated. Make sure that the joystick is in
the center position before switching on the control system.
6 Bars flash
Battery charger connected.
5 Bars flash
Right-hand motor wiring fault.
4 Bars flash
Right-hand motor disconnected.
3 Bars flash
Left-hand motor wiring fault.
2 Bars flash
Left-hand motor disconnected.
1 Bar flashes
Low battery voltage.