23

Appendix

Specification

Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura-

tion measuring instrument

• Name:

Oxygen Saturation Monitor

• Model name: PULSOX-300

Functions:

• Measuring range: SpO

2

:

0 to 100% SpO

2

Pulse rate: 30 to 230 bpm

• Accuracy:

SpO

2

:

±2% SpO

2

(70 to 100% range, 1S.D.)

Pulse rate: ±2 bpm (30 to 100 range)

or ±2% of value (100 to 230 range)

PULSOX-300 has been clinically validated for use on adults.

Display

Display type: Liquid crystal display

Oxygen saturation (SpO

2

)

Pulse rate number

Pulse level meter (5 steps)

Battery indication

Error messages indication

Warning functions

SpO

2

: Low SpO

2

warning (flashing display)

Battery indication (flashing when the battery is nealy empty)

Error message indication

Operating conditions

temperature/humidity range

0 to 40°C (32 to 104°F); 30 to 85% relative humidity with no condensation

atmospheric pressure/altitude range

700 to 1060hPa (altitude: –400 to 3000m)

Storage conditions

temperature/humidity range

–10 to 60°C (14 to 140°F); 10 to 95% relative humidity with no condensation

atmospheric pressure/altitude range

700 to 1060hPa (altitude: –400 to 3000m)

Power

1 AAA-size battery: 1.5V 100mW (Service life under continual usage:

approx. 16 hours with alkaline battery)

Dimensions (W × H × D)

68 Ч 58 Ч 15 mm

Weight

56g (including battery/excluding wrist band and probes)

Usable life

6 years [Verified by KONICA MINOLTA SENSING, INC. (based on own data)]