Calculation Method for Displayed Values

<Pulse level meter>

The pulse level meter displays the transmitted light amount ratio calculated by the

following equation. The pulse level is indicated in eight levels (0 to 15%, full scale

if over 10%).

<Oxygen saturation (SpO

2

)>

Updates the moving average of the oxygen saturation over the last 12 seconds at

one-second intervals.

<Pulse rate (P.R.)>

Updates the moving average of the last 8 pulse rates at one-second intervals.

Pulse level (%) = ------------------------------------------------------------------------------------------------------------------ x 100

Transmitted light amount that does not vary due to the pulse

Transmitted light amount that varies due to the pulse