Message display

Host must return
all hyper’s disc

scsiID=7 TermOFF
dev-ID=

E88, E89,E99, etc.

¶ Memory Failure

¶ Error

Possible cause

There are some discs that from
the hyper magazine remaining in
drives.

¶ SCSI cable has not been

connected properly.

¶ SCSI ID or termination

settings have not been
correctly specified.

¶ The total length of the SCSI

cables is too long.

¶ Operator has forgotten to

reboot the host computer.

¶ The power cable for the drive

in question has not been
connected.

¶ Drives have been inserted in

rear bays #9 through #16
without additional power
supply unit.

¶ The disc has been inserted

upside down.

¶ The disc is soiled or

scratched.

A mechanical error has occurred
which has made the changer
unable to continue in operation.

The changer SCSI termination
switch is set to OFF or the SCSI
cable is not properly connected.

Magazine's memory data is not
being read correctly.

Description of problem

Hyper magazine cannot be ejected
in Hyper submode.

There exists a drive or drives which
fail to be recognized by the host
computer even though they have
been recognized in Config submode.

There exists a drive which fails to be
recognized by the host computer,
and the SCSI ID of the drive is set to
‘7’ when checked in Config
submode.

Unable to properly read a disc which
has been inserted in a drive.

An error code is displayed on the
LCD message window and the
backlight flashes and the buzzer is
sounded.

The LCD message window does not
come on when the power is turned
on.
Pressing the operation keys has no
effect.

Buzzer sounds when magazine is
loaded.

Recommended action

Return to user mode, issue REZERO
command from the host computer
requesting that all discs be returned,
and try to perform the operation again
after all discs have been returned.

¶ Make sure that the SCSI cables have

been connected properly.

¶ Check to see if there is any device on

the same bus with the same SCSI ID
and to see that the SCSI chain has
been properly terminated.

¶ Re-configure the SCSI chain so that

the total length of all the cables does
not exceed the maximum allowable
length.

¶ Reboot the host computer.

¶ Open the rear access door and check

to see if the power cable has been
properly connected.

¶ The power supplied by the power

supply included as standard
equipment is sufficient only for
operating drives in bays #1 through
#8. Reinstall the drive in one of these
bays or install the additional power
supply unit.

¶ Check to see if the disc has been

inserted upside down.

¶ Check to see if the disc is soiled or

scratched, and wipe away any soil or
replace the disc.

In system administrator mode, open the
access door and take whatever action is
required in order to eliminate the error
condition.

¶ When installing is not finished, turn

the SCSI termination switch to ON
and turn the power OFF and ON.

¶ If the SCSI cable is not properly

connected, turn off the power. And
after connection the cable properly,
turn on the power again.

¶ Remove and reinsert magazine.

¶ If the magazine data still can't be read

properly, perform a "PROBE" of the
magazine, ten open and close the
door.