INSPECTION AND CLEANING
Perform the following inspection and cleaning procedures periodically as a
preventive maintenance measure to keep the Polaris System in optimal
condition resulting in trouble-free operation producing crisp, clear images .

Inspection - Routinely inspect each component of the Polaris System for
possible defects as follows:

1 . Camera Handpiece -

a . Check overall handpiece for chips, cracks or other irregularities .
b . Check the lens window for debris or spots .
c . Check the connector socket for damage .

2 . Control Module -

a . Check for broken indicator lamp and switch .
b . Check all connector sockets for damage, loose or missing hardware .

3 . Cables and Connectors -

a . Check cables for damaged or deteriorated insulation kinking or

twisting .

b . Check connectors for loose, bent or missing pins .
c . Check that both quick-disconnect plug ends snap closed to snugly

secure the Handpiece Cable to the Camera Handpiece and Control

Module connectors .

CAUTION: Always turn Off power and unplug the Polaris System components before

performing cleaning procedures.

Do not attempt any internal service of Polaris System components. Contact your local autho-
rized Air Techniques dealer for service. Failure to heed this warning may result in equipment
damage and voiding the warranty.

MAINTENANCE

Do not spray cleaning liquids or disinfectants directly on the camera handpiece. Use

care not to allow liquids to run into internal circuitry.
Do not wipe the surfaces using benzine, thinner, etc. as this may degrade the finish.
Do not autoclave the Camera handpiece.

2

The camera cover must be used for only one patient and disposed

of properly in accordance with local code.

Cleaning - Clean each component of the Polaris System as needed by
performing the following procedures .

1 . Control Module-Use a non-corrosive surface disinfectant such as Birex® se .
2 . Camera Handpiece-Following every patient use, remove the camera cover

from the Camera Handpiece and dispose in accordance with local regula-

tions . The Camera Handpiece can be wiped clean and disinfected with

any non-corrosive surface disinfectant such as Birex® se .

3 . Camera Window Lens-Remove debris or spots from lens window by

cleaning with alcohol and cotton swabs .