NI WLS/ENET-9163 Carrier User Guide and Specifications

Specifications

These specifications are typical at 25 °C unless otherwise noted.

For C Series I/O module specifications, refer to the documentation
included with the modules.

Note

These specifications are for the NI WLS/ENET-9163 carrier only, unless otherwise

noted.

Analog Input

Input FIFO size ...................................... 4095 samples >16 bit

8191 samples

≤16 bit

Sample rate

1

NI WLS/ENET-9163 carrier........... 5 MS/s (multi-channel,

aggregate), maximum

With NI WLS/ENET-9215 ............. 100 kS/s, maximum

Timing accuracy

2

................................... 50 ppm of sample rate

Timing resolution

2

................................. 50 ns

Number of channels supported .............. Determined by the C Series

I/O module

Digital Triggers

Static Characteristics

Number of terminals .............................. 2 bi-directional, individually

settable

Pull-down resistor .................................. 49.9 k

Ω ±0.5%

Input voltage protection

3

........................ ±20 V on each pin

Power–on state ....................................... Input

Required minimum input pulse width.... 100 ns

1

Performance dependent on type of installed C Series I/O modules and number of channels in the task.

2

Does not include group delay. Refer to C Series I/O module documentation for more information.

3

Stresses beyond those listed under Input voltage protection may cause permanent damage to the device.