Appendix A

Specifications

NI PXI-8184/8185 User Manual

A-6

ni.com

Caution

Clean the PXI-8184 with a soft nonmetallic brush. Make sure that the device is

completely dry and free from contaminants before returning it to service.

Storage Environment

Ambient temperature range ....................–20 to 65 °C (Tested in

accordance with IEC-60068-2-1
and IEC-60068-2-2.)

Relative humidity range..........................5% to 95% noncondensing

(Tested in accordance with
IEC-60068-2-56.)

Shock and Vibration

Operational shock ...................................30 g peak, half-sine, 11 ms pulse

(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)

Random vibration

Operating .........................................5 to 500 Hz, 0.3 g

rms

(with solid-state hard drive)

Nonoperating ...................................5 to 500 Hz, 2.4 g

rms

(Tested in accordance with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-PRF-28800F, Class B.)

Note

For EMC compliance, operate this device with shielded cabling. In addition, all

covers and filler panels must be installed. Refer to the Declaration of Conformity (DoC)
for this product for any additional regulatory compliance information. To obtain the DoC
for this product, visit

ni.com/certification

, search by model number or product line,

and click the appropriate link in the Certification column.

Safety

This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1

CAN/CSA C22.2 No. 61010-1