Calibration, Bus interface, Power requirement – National Instruments NI 7831R User Manual

Page 50

Advertising
background image

Appendix A

Specifications

NI 7831R User Manual

A-6

ni.com

Phase locked to PXI 10 MHz
Clock(NI PXI-7831R only) ....................Adds 350 ps jitter, 300 ps skew

Additional frequency dependent jitter

40 MHz ............................................None

80 MHz ............................................400 ps

120 MHz..........................................720 ps

160 MHz..........................................710 ps

200 MHz..........................................700 ps

Calibration

Recommended warm-up time.................15 minutes

Calibration interval .................................1 year

Onboard calibration reference

DC level...........................................5.000 V (

±3.5 mV)

(actual value stored
in Flash memory)

Temperature coefficient

...................±5 ppm/°C max

Long-term stability

..........................±20 ppm/

Note

Refer to Calibration Certificates at

ni.com/calibration

to generate a

calibration certificate for the NI 7831R.

Bus Interface

PXI (NI PXI-7831R only) ......................Master, slave

PCI (NI PCI-7831R only).......................Master, slave

Power Requirement

+5 VDC (

±5%) .......................................450 mA (typ), 700 mA (max)

(does not include current drawn
from the +5 V line on the
I/O connectors)

+3.3 VDC (

±5%) ....................................335 mA (typ), 730 mA (max)

(does not include current sourced
by the digital outputs. To
calculate the total current sourced

1,000 h