Technical specifications, Humidifier performance, Symbols which appear on the device – ResMed H4i User Manual

Page 11: En gli sh

Advertising
background image

9

Technical Specifications

En

gli

sh

Technical Specifications

Humidifier Performance

The following settings have been tested at 22°C, 40% RH ambient conditions:

Note: The manufacturer reserves the right to change these specifications without
notice.

Symbols which appear on the device

Attention, consult accompanying documents;

Type CF equipment;

Class II equipment;

Hot surface;

Do not touch;

Maximum water level.

Maximum heater plate temperature

185°F (85°C)

Temperature cut-out

199°F (93°C)

Maximum gas temperature

106°F (41°C)

Dimensions (H x W x D)

Docking station and water chamber: 5.0” x 6.7” x 8.3” (130 mm x
170 mm x 212 mm)

Weight

Docking station and unfilled water chamber: 1.6 lb (732g)

Water capacity

To maximum fill line: 390 mL

Power supply

Input range for ResMed S8 flow generator with H4i: 100–240V,
50/60Hz; 110V, 400Hz; 2.5A <140 VA (110 W) (max. power
consumption) Instantaneous peak power consumption <340 VA

*

Maximum heater element power: 85 W

*

Actual average power consumption will vary depending on factors such as temperature setting, set
pressure, the use of accessories, height above sea level and ambient temperature.

Docking station

Flame retardant engineering thermoplastic, plated cast aluminium

Water chamber

Injection molded plastic, stainless steel and silicon seal

Operating temperature

+41°F to +95°F (+5°C to +35°C)

Operating humidity

10–95% non-condensing

Storage and transport temperature

-4°F to +140°F (-20°C to +60°C)

Storage and transport humidity

10–95% non-condensing

IEC 60601-1 classification

Class II (double insulation), Type CF

Electromagnetic compatibility

Please refer to your flow generator user's manual for details

CPAP Mask Pressure
cm H

2

O

RH Output %

Setting 3

Setting 6

Nominal System Output AH

*

, BTPS

Setting 3

Setting 6

4

>95

>95

16

20

10

>85

>95

14

18

20

>75

>85

13

16

*

AH- Absolute Humidity in mg/L.

BTPS- Body Temperature Pressure Saturated.