Weekly, Cover and cover seal, Docking station – ResMed H4i User Manual

Page 8

Advertising
background image

6

Weekly

Inspect the water chamber for wear and deterioration. Replace the water chamber if
any component has cracked, become cloudy or pitted. Replace immediately if there
is any leaking from the water chamber.
White powder or deposits in the water chamber can be cleaned using a solution of
one part vinegar to 10 parts water, as follows:

1

Fill the cleaning plug to the rim with vinegar and pour into empty water chamber.

2

Add clean water to the water chamber until the solution reaches the maximum water
level mark.

3

Let the solution soak for 10 minutes.

4

Discard the solution and rinse the water chamber with clean water.

Cover and Cover Seal

Inspect the cover and cover seal weekly for wear and deterioration. Replace any
damaged parts. When required, wash the cover and cover seal in warm water, using
a mild detergent. Rinse them thoroughly with clean water and allow it to dry away
from direct sunlight.

Docking Station

Periodically wipe the outside of the docking station (including the connector seal) with
a damp cloth.

!

CAUTION

Do not use any of the following to clean any part of the H4i as these
solutions may cause damage and reduce the life of the product:
– bleach, chlorine-, alcohol- or ammonia-based solutions
– moisturizing, antibacterial or glycerine based soaps
– water softening and unapproved descaling agents.

Do not wash the water chamber in a dishwasher.

Advertising