ResMed S8 User Manual

Page 12

Advertising
background image

7

E

NG

LISH

M

EDICAL

I

NFORMATION

If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
If oxygen has been left on, turn off the flow generator, then wait 30 minutes before
turning on the flow generator again.
Explanation: When the CPAP device is not in operation and the oxygen flow is
left on, oxygen delivered into the air delivery tubing may accumulate within the
CPAP machine enclosure and create a risk of fire. This applies to most types of
CPAP machines.

Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.

Always ensure airflow is being generated by the device before the oxygen
supply is turned on.

Always turn the oxygen supply off before stopping the airflow from the device.

Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will
vary, depending on where the oxygen is introduced, the pressure settings, patient breathing
pattern, mask selection and leak rate.

Do not use the

S8 E

SCAPE

if there are obvious external defects or unexplained

changes in performance.

Do not open the

S8 E

SCAPE

case. There are no user serviceable parts inside.

Repairs and internal servicing should only be performed by an authorised service
agent.

Explosion hazard—do not use in the vicinity of flammable anaesthetics.

C

AUTIONS

At low pressures, the flow through the exhalation ports of your mask may not
clear all exhaled gas from the tubing. Some rebreathing may occur.

The airflow for breathing produced by this device can be as much as 6°C higher
than the temperature of the room. Caution should be exercised if the room
temperature is warmer than 32°C.

When AC mains power (100–240V AC) is not available, always use a ResMed
DC-12 converter. (The DC-12 converter is available as an optional accessory. It is
not supplied with all models.)

If supplied, the Data Card module is designed for use with specific ResMed
devices.

Only use the Data Card module (if supplied) as specified in this manual.

Note: The above are general warnings and cautions. Specific warnings, cautions and notes
appear with the relevant instructions in the manual.

338223r1.book Page 7 Friday, June 3, 2005 4:16 PM

Advertising