If oxygen is used with the ventilator, the oxygen flow must be turned off
when the ventilator is not operating.
Explanation: When the ventilator is not in operation, and the oxygen flow
is left on, oxygen delivered into the ventilator tubing may accumulate within
the ventilator enclosure. Oxygen accumulated in the ventilator enclosure
will create a risk of fire. This warning applies to most types of ventilators.

Please refrain from smoking while oxygen is in use.

About the Mask

The mask system does not contain latex, PVC or DEHP materials. If the
patient has ANY reaction to any part of the mask system, discontinue use
and investigate the possible causes.

ResMed has tinted the Ultra Mirage NV Full Face Mask elbow blue to
identify it as a non-vented mask for use with active exhaust non-invasive
ventilator systems only.

The mask has a standard female 22 mm tapered fitting (ISO5356-1) for
external connection to the ventilator.

Refer to the section “Technical Specifications” of this guide for information
about mask leak, dead space and resistance.

CAUTION

The mask must be cleaned and disinfected/sterilized if it is used between
patients. Refer to the section “Cleaning the Mask between Patients” for
details.

Parts of the Mask

See section

A on the illustrations sheet.

Fitting the Mask

Note: Refer to the relevant ventilator system operating instructions for detailed
information about connecting the mask.

See section

B on the illustrations sheet.

The Ultra Mirage NV Full Face Mask is supplied fully assembled.
1 Move the forehead support tab to Position 3. (Refer illustration B-1 for

location of Positon 3.)

2 Disconnect one lower headgear clip. Place the mask on the face and pull the

headgear over the patient’s head

(B-2). Ensure that the upper straps pass

across the forehead and the lower straps below the ears.

3 Reattach the headgear clip to the mask frame.
4 Gently tighten the upper headgear straps (B-3).