Technical specifications, Symbols which appear on the device, En gli sh – ResMed VPAP AUTO 25 User Manual

Page 18

Advertising
background image

15

Technical Specifications

En

gli

sh

Technical Specifications

Note: The manufacturer reserves the right to change these specifications without notice.

Symbols Which Appear On The Device

Attention, consult accompanying documents;

Drip proof;

Type CF equipment;

Dangerous voltage;

Class II equipment;

Start/Stop;

Manufacturer.

Operating pressure range

3 to 25 cm H

2

O

Maximum single fault steady
state pressure

40 cm H

2

O

Pressure measurement
tolerance

±0.5 cm H

2

O ± 4% of the measured reading

Flow measurement tolerance

±0.1 or 20% of reading, whichever is greater

VAuto mode

4 to 25 cm H

2

O (measured at the mask); Min EPAP: 4 cm H

2

O;

Max IPAP: 25 cm H

2

O; Pressure Support: 0 to 10 cm H

2

O

S, ST and T modes

IPAP: 4 to 25 cm H

2

O (measured at the mask); EPAP: 3 to 25 cm H

2

O

(measured at the mask); Pressure Support: 0 to 22 cm H

2

O

CPAP mode

4 to 20 cm H

2

O (measured at the mask)

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:

Sound pressure level

26 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1: 2002
28 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007

Sound power level

36 dBA with uncertainty of 2 dBA as measured according to
ISO 17510-1:2007

Dimensions (L x W x H)

4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)

Weight

2.9 lb (1.3 kg)

Power supply

Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)

Housing construction

Flame retardant engineering thermoplastic

Operating temperature

41

º

F to 95

º

F (+5

º

C to +35

º

C)

Operating humidity

10–95% non-condensing

Storage and transport
temperature

-4

º

F to 140

º

F (-20

º

C to +60

º

C)

Storage and transport humidity

10–95% non-condensing

Operating altitude

Sea level to 8,500’ (2,600 m)

Supplemental oxygen

Recommended maximum supplemental oxygen flow:
4 L/min (VAuto mode); 15 L/min (CPAP, S, ST and T modes)

Electromagnetic compatibility

Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments

Air filter

Two-layered, powder-bonded, polyester non-woven fiber

Air tubing

Flexible plastic, 1 x 6’6” (2 m)

Air outlet

The 22 mm conical air outlet complies with ISO 5356-1

IEC 60601-1 classification

Class II (double insulation), Type CF