Supplemental oxygen, Using the vpap auto 25, vpap st & vpap s, Setting up your vpap – ResMed VPAP AUTO 25 User Manual

Page 6

Advertising
background image

3

Using the VPAP Auto 25, VPAP ST & VPAP S

En

gli

sh

Supplemental Oxygen

The VPAP Auto 25, VPAP ST and VPAP S are designed to be compatible with the
following levels of supplemental oxygen usage:

Up to 4 L/min in VAuto mode

Up to 15 L/min in CPAP, S, ST and T modes.

At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on the pressure settings, patient breathing pattern, mask
selection, and the leak rate.

!

WARNING

Always use the ResMed Oxygen Connector Port when adding supplemental
oxygen at the flow generator outlet.

Using the VPAP Auto 25, VPAP ST & VPAP S

Setting up your VPAP

1

Connect the power cord to the socket at the rear of your VPAP.

2

Plug the other end of the power cord into the power outlet.

3

Connect one end of the air tubing firmly onto the air outlet.

4

Connect the assembled mask system to the free end of air tubing.
For information on assembling your mask, see your mask manual.

!

WARNING

Make sure the power cord and plug are in good condition and the
equipment is not damaged.

Only ResMed air tubing should be used with the device. A different type of
air tubing may alter the pressure you actually receive, reducing the
effectiveness of your treatment.

Blocking the hose and/or air inlet of the device while in operation could lead
to overheating of the device.

1

2

3

4

Advertising
This manual is related to the following products: