SPECIFICATIONS

(Typical @ 25C and rated supply voltage unless otherwise specified)


STANDARD INPUTS

* 4-20ma 10-50ma 0-5Vdc 0-10Vdc

* Thermocouple inputs: J,K,T,E,R,S,B

* Cold junction compensation error: +/- 1C (10C to 40C)

* Open input indication: HELP displayed

* Temperature displayable in Degrees C or F

* Non standard inputs available --- consult factory.


ANALOG TO DIGITAL CONVERSION

* 4-1/2 Digit (20,000) Count) A/D Converter

* Dual slope integrating converter with 7 conversions /sec. (typical) rate.


SCAN RATE

* Fixed: two channels per second


DISPLAY

* Red seven segment displays--0.39" (10mm) digit height

* Over range indication: HELP

* Display test: Briefly displays 8.8.8.8.8.8.8 on power up.

* Seven digit display


SCALE/OFFSET

* Scale programmable from 1-30000 units

* Offset programmable from 0-99.99 MA (For MA Input)

* Offset programmable from 0-9.999 MV (For Voltage Input)

* Programmable decimal point.


RELIABILITY/ACCURACY
*Calibration: NIST traceable (for thermocouples)
*Temperature resolution: 1C/1F
* Warranty: 1 Year

* Recalibration recommended at 12 month interval


POWER REQUIREMENT
*

120Vac

--- 60 HZ

(Standard)

*

240Vac

--- 50 HZ

(optional)

*

15VDC (optional)


OUTPUT (optional)

* Open collector outputs -- 100ma maximum sink, limited to 50ma per output

* Relays: Single pole single throw, 1 Amp @ 28Vdc or 0.5 Amp @ 120Vac resistive
*Output termination: Euro-style plug-in connector