10.2.4- JUMPERS CONFIGURATION

Function

Close Jumper

220

terminator resistor for the RS422

JP1 (By default opened)

switch debounce of 5 Hz for the STOP line

JP2* (By default closed)

switch debounce of 5 Hz for the START line

JP3* (By default closed)

switch debounce of 5 Hz for the RESET line

JP4* (By default closed)

10.2.5- SERIAL DATA OUTPUT CONFIGURATION

* = Factory set-up

* = Data output format only when the colon is lighted.
** = Data output format only when the colon is NOT lighted.

H = Character from "0" to "9" which represent one number of hours.
M = Character from "0" to "9" which represent one number of minutes.
S = Character from "0" to "9" which represent one number of seconds.

Character "." = 02E hex.

Character "-" = 02D hex.

Character "(NUL)" = 0 hex.

Character "(CR)" = 0D hex.

Jumper Application

A

Jumper position

B

JP5

for STOP line

1 K

pull-up *

1 K

pull-down

JP6

for START line

falling edge/low level active*

rising edge/high level active

JP7

for STOP line

falling edge/low level active*

rising edge/high level active

JP8

for RESET line

falling edge/low level active*

rising edge/high level active

JP9

for START line

1 K

pull-up*

1 K

pull-down

JP10

for RESET line

1 K

pull-up*

1 K

pull-down

Data output format
with DIP switch 7

with DIP switch 7

in position ON

in position OFF

(NEWPORT mode)

Standard mode

C2

HHM.MS.S(CR)

HH.MM.SS(CR)

--(NUL)--(NUL)--(CR)

C1

HHM.M(CR)

HH.MM(CR) *

--(NUL)--(CR)

HH(NUL)MM(CR)**

C0

MMS.S(CR)

MM.SS(CR)

--(NUL)--(CR)

Instrument
reference

Transmission during
time adjust or in
alarm status