Omega CDE-45P

Conductivity Sensor

Part 3 –Maintenance & Troubleshooting

O&M Manual

- 10 -


3.2

Troubleshooting

The first step in resolving any measurement problem is to
determine whether the trouble lies in the sensor or the
transmitter. Since measurement problems can often be
traced to surface deposits coating the electrodes, cleaning
the sensor using the method outlined in Section 3.1 should
always be the first step in any troubleshooting.

If the sensor cannot be calibrated after cleaning, perform the
following test. A multimeter will be needed.

1.

Disconnect the sensor from the transmitter or junction

box.

2.

Using a multimeter, verify continuity between

electrodes indicated below and the corresponding

wire colors in the cable (red, blue, green and white,

only).

3.

Verify that the sensor’s temperature element (Pt1000

RTD) is functioning properly by measuring the

resistance between the sensor’s yellow and black

wires. The nominal resistance value at 25 ºC is 1097

ohms. Use the following table as a guide to the

approximate resistance value: