OMEGA CDE-45P

Conductivity Sensor

Part 2– Installation

O&M Manual

- 5 -

Part 2 – Installation


2.1

General

The CDE-45P Conductivity Sensor is designed for
industrial

and

municipal

process

applications.

Mounting options include flow-through, submersion,
insertion (special hardware required), or integral
mount to the CDTX-45 Conductivity Monitor/Analyzer.
The sensor-to-analyzer distance must not exceed 60
feet (18.2 meters).

Calibrate the sensor before placing it into the process.
See Model CDTX-45 Monitor/Analyzer Instruction
Manual for detailed calibration instructions.

The sensor comes with a removable guard that
surrounds the electrode face. This guard minimizes
interference effects in tight locations where the sensor
face is close to surrounding objects. If it is removed,
take care to leave at least 1 in

3

of space in front of the

electrodes. If the guard is to be used, the sensor
must be calibrated with it in place, since the guard
affects the sensor cell constant.