2 troubleshooting guide 35 – Omega FMA 4000 User Manual

Page 39

Advertising
background image

9.2 Troubleshooting

Guide

35

NO.

INDICATION

LIKELY REASON

SOLUTION

1

No zero reading after
15 min. warm up time
and no flow condition.

Embedded temperature
has been changed.

Perform Auto Zero Procedure (see section
5.3.6 “Zero Calibration”).

2

Status LED indicator
and LCD Display
remains blank when
unit is powered up. No
response when flow is
introduced from analog
outputs 0-5 Vdc or
4-20 mA.

Power supply is bad or
polarity is reversed.

Measure voltage on pins 7 and 5 of the 15
pin D-connector. If voltage is out of
specified range, then replace power supply
with a new one. If polarity is reversed
(reading is negative) make correct
connection.

PC board is defective.

Return FMA 4000 to factory for repair.

3

LCD Display reading or
/ and analog output
0-5 Vdc signal fluctuate
in wide range during
flow measurement.

Output 0-5 Vdc signal
(pins 2–1 of the
D-connector) is shorted
on the GND or
overloaded.

Check external connections to pin 2 – 1, of
the D-connector. Make sure the load
resistance is more than 1000 Ohm.

4

LCD Display reading
does correspond to the
correct flow range, but
0-5 Vdc output signal
does not change
(always the same read
ing or around zero).

Output 0-5 Vdc
schematic is burned
out or damaged.

Return FMA 4000 to factory for repair.

Analog flow output
scale and offset
variable are corrupted.

Restore original EEPROM scale and offset
variable or perform analog output
recalibration (see section 7.3).

5

LCD Display reading
and 0-5 Vdc output
voltage do correspond
to the correct flow
range, but 4-20 mA
output signal does not
change (always the
same or reading
around 4.0 mA).

External loop is open or
load resistance more
than 500 Ohm.

Check external connections to pins 2 and
15 of the D-connector. Make sure the loop
resistance is less than 500 Ohm.

Output 4-20 mA
schematic is burned
out or damaged.

Return FMA 4000 to factory for repair.

6

Calibration is off (more
than ±1.0 % F.S.).

FMA 4000 has initial

zero shift.

Shut off the flow of gas into the FMA 4000
(ensure gas source is disconnected and no
seepage or leak occurs into the meter).
Wait for 15 min. with no flow condition and
perform Auto Zero calibration Procedure
(see section 5.3.7 “Zero Calibration”).

7

LCD Display reading is
above maximum flow
range and output volt
age 0-5 Vdc signal is
more than 5.0 Vdc
when gas flows
through the FMA 4000.

Sensor under
swamping conditions
(flow is more than 10%
above maximum flow
rate for particular
FMA 4000).

Lower the flow through FMA 4000 within
calibrated range or shut down the flow
completely. The swamping condition will
end automatically.

PC board is defective.

Return FMA 4000 to factory for repair.

Advertising