Changing the air filter – Omron Healthcare COMPAIV XLT NE-C25 User Manual

Page 19

Advertising
background image

19

Change the air filter every 30 days even if the air filter does not appear
dirty. If the air filter appears dirty, or if water or medication has spilled on
the air filter replace with a new air filter immediately.

1. Pull the air filter cover to remove it from the front side of

the compressor.

2. Remove the dirty filter with a small

pointed object.

CAUTION

Do not attempt to wash or clean the air filter. Damp

air filters can cause blockages. Do not substitute
cotton or any other material for the air filter.

CAUTION

Wash the air filter cover regularly to prevent any
blockage in the cover. Do not boil. Make sure the
cover is dry before inserting the new air filter.

CAUTION

Do not operate without air filter.

3. Insert a new air filter into the air filter cover.

CAUTION

Before inserting the new air filter make sure the air filter is clean and free of
dust. Use only the Omron air filter designed for this device.

4. Put the air filter cover back on the compressor.

CHANGING THE AIR FILTER

272145 NE-C25-all 11/5/07 2:28 PM Page 19

Advertising