Multiple on-board cryopump connections, Caution, Helium line connections – Olicom 9600 User Manual

Page 34: Multiple on-board cryopump connections -11, Helium line connections -11, Ti cryogenics

Advertising
background image

9600 Compressor Installation, Operation, and Maintenance

P/N 8040540

3-11

C

HELIX TECHNOLOGY CORPORATION

-

TI

CRYOGENICS

Multiple On-Board Cryopump Connections

CAUTION

Make sure the Compressor power is OFF before making any
connections to the rear panel.

Helium Line Connections

CAUTION

The use of several compressors on a single manifold feeding a common
supply header and a common return header requires special
precautions. Contact CTI-CRYOGENICS for a review of the intended
installation and for specific technical instructions. The use of a 9600
compressor on a manifold with other CTI-CRYOGENICS compressor
models requires a reduction of the helium charge pressure to 200 - 210
psig charge pressure to avoid helium safety valves from inadvertently
venting. Refer to Section 4 - Operation and Section 5 - Maintenance
for more information.

1. Connect the Supply and Return lines to the 9600 Compressor as

described in Connecting/Disconnecting Helium Flex Lines
within this section.

2. Connect the Gas Return Line to the customer supplied helium

manifold and then to the GAS RETURN connector on the
On-Board Cryopump.

3. Connect the Gas Supply Line to the customer supplied helium

manifold and then to the GAS SUPPLY connector on the On-Board
Cryopump.