Teledyne 3020T User Manual

Page 44

Advertising
background image

4 Operation

Model 3020T

4-8

Teledyne Analytical Instruments

A A A A A

Retype PWD To Verify

Wait a moment for the entry (

<ENT>) screen. You will be given

clearance to proceed.

A A A A A

<ENT> TO Proceed

ENTER the letters of your new password. Your password will be stored

in the microprocessor and the system will immediately switch to the ANA-
LYZE screen, and you now have access to all instrument functions.

If all alarms are defeated, the ANALYZE screen appears as:

0.0 ppm Anlz

Range: 0 — 100

If an alarm is tripped, the second line will change to show which alarm

it is:

0.0 ppm Anlz

AL—1

NOTE:If you log off the system using the LOGOUT function in the

MAIN MENU

, you will now be required to re-enter the pass-

word to gain access to

SPAN, ZERO, ALARM,

and

RANGE

functions.

4.5

The LOGOUT Function

By entering

LOGOUT

, you effectively log off the instrument, leaving

the system protected against tampering until the password is reentered. To
log out, scroll to place the

LOGOUT

function between the arrows, and

ENTER to log out. The screen will display the message:

Protected Until

Password Reentered

4.6

The VERSION Screen

Scroll through the MAIN MENU to

VERSION

, and ENTER. The

screen displays the manufacturer, model, and software version information.

Advertising
This manual is related to the following products: